Chapter 7

Digital meters and electronic counters

A digital readout adds quantisation and count specifications to the analogue front-end errors. Identify the input function, selected range and least-significant digit before computing accuracy. Counter resolution depends on which signal opens the gate.

7.1 Digits, counts and resolution

A conventional d+12d+\tfrac12 digit display has dd unrestricted decimal digits and a leading digit limited to 0 or 1. Modern count ratings can differ; the stated maximum display takes precedence.

Conventional displayMaximum countExample selected rangeLeast count
3½ digits1999Maximum 1.999 V1 mV
4½ digits19999Maximum 19.999 V1 mV
4 full digits9999Maximum 9.999 V1 mV

7.2 DMM accuracy and input functions

Voltage mode uses a high input resistance; current mode measures the drop across a shunt; resistance mode measures voltage produced by a known test current. AC mode may be average responding, AC true RMS, or AC+DC true RMS.

7.3 Integrating DVM behaviour

Converter architectures and sample-and-hold operation are treated in the Digital Logic and Electronics handbook. For measurement problems, retain the integration-time and count relations and their noise-rejection conditions.

7.4 Frequency and period counters

Direct counting counts unknown cycles during a known gate interval. Reciprocal counting measures the duration of a known number of unknown cycles using a fast reference clock.

7.5 Electronic phase and trigger measurements

Compare corresponding edges or same-slope zero crossings of equal-frequency signals. Declare which channel is delayed. Noise near the trigger threshold changes the measured crossing time.

Last-minute recall

  • A half digit has a restricted leading value; use counts when supplied.
  • One count is the selected range's least-significant increment.
  • Reading-percent and range-percent errors are different terms.
  • DMM resistance mode supplies a test current.
  • Dual-slope conversion compares two volt-second areas.
  • Integral-cycle averaging rejects periodic interference ideally.
  • Longer direct-count gates improve frequency resolution.
  • Reciprocal counting improves low-frequency measurement.
  • Trigger noise divided by signal slope gives timing uncertainty.