Chapter 8

Oscilloscopes

An oscilloscope measures voltage versus time or one voltage versus another. Include probe attenuation, input coupling and bandwidth in the measurement model. A stable trace does not by itself prove that the waveform is sampled without aliasing.

8.1 Controls and voltage–time readings

Vertical gain sets volts/division; the timebase sets seconds/division. Trigger level and slope select a reproducible starting point. DC coupling retains DC; AC coupling blocks DC and distorts sufficiently slow variations. The delay line in an analogue CRO lets the sweep start before the signal reaches the vertical plates. Dual-trace instruments switch one beam between channels; dual-beam instruments have separate beams.

8.2 Equal-frequency Lissajous phase

In XY mode let x=Asin⁡ωtx=A\sin\omega t and y=Bsin⁡(ωt+ψ)y=B\sin(\omega t+\psi). Here ψ\psi is the phase of the Y signal relative to X; it is distinct from the voltage-minus-current convention until channels are identified.

Original XY traces for equal displayed amplitudes: zero, 45-degree and 90-degree phase offsets.Original XY traces for equal displayed amplitudes: zero, 45-degree and 90-degree phase offsets.

8.3 Lissajous frequency ratio

A stationary closed trace requires a stable rational frequency ratio. Count contacts at the extreme top and extreme right over the full closed trace, without double-counting a retraced point. Avoid degenerate phases where distinct contacts merge.

8.4 Probe compensation and bandwidth

A passive attenuating probe reduces resistive loading but adds capacitance. Matching the probe and input time constants makes its attenuation approximately frequency-independent in the lumped model. A mismatched probe rounds or overshoots a square-wave edge.

8.5 CRT deflection and digital acquisition

In a CRT, accelerating voltage sets electron speed and deflection plates supply a transverse field. In a digital storage oscilloscope, sampling and memory replace the screen's persistence as the record of the event.

Last-minute recall

  • Apply the probe factor exactly once.
  • Peak-to-peak to RMS conversion depends on waveform shape.
  • AC coupling removes DC and can distort slow waveforms.
  • A Lissajous ellipse gives phase magnitude with sign ambiguity.
  • Count tangent contacts with a defined axis convention.
  • Match probe and scope RC time constants.
  • The 0.35 rise-time rule depends on frequency-response shape.
  • Sample rate, record length and analogue bandwidth are separate limits.