Oscilloscopes
An oscilloscope measures voltage versus time or one voltage versus another. Include probe attenuation, input coupling and bandwidth in the measurement model. A stable trace does not by itself prove that the waveform is sampled without aliasing.
8.1 Controls and voltage–time readings
Vertical gain sets volts/division; the timebase sets seconds/division. Trigger level and slope select a reproducible starting point. DC coupling retains DC; AC coupling blocks DC and distorts sufficiently slow variations. The delay line in an analogue CRO lets the sweep start before the signal reaches the vertical plates. Dual-trace instruments switch one beam between channels; dual-beam instruments have separate beams.
8.2 Equal-frequency Lissajous phase
In XY mode let and . Here is the phase of the Y signal relative to X; it is distinct from the voltage-minus-current convention until channels are identified.
Original XY traces for equal displayed amplitudes: zero, 45-degree and 90-degree phase offsets.
8.3 Lissajous frequency ratio
A stationary closed trace requires a stable rational frequency ratio. Count contacts at the extreme top and extreme right over the full closed trace, without double-counting a retraced point. Avoid degenerate phases where distinct contacts merge.
8.4 Probe compensation and bandwidth
A passive attenuating probe reduces resistive loading but adds capacitance. Matching the probe and input time constants makes its attenuation approximately frequency-independent in the lumped model. A mismatched probe rounds or overshoots a square-wave edge.
8.5 CRT deflection and digital acquisition
In a CRT, accelerating voltage sets electron speed and deflection plates supply a transverse field. In a digital storage oscilloscope, sampling and memory replace the screen's persistence as the record of the event.
Last-minute recall
- Apply the probe factor exactly once.
- Peak-to-peak to RMS conversion depends on waveform shape.
- AC coupling removes DC and can distort slow waveforms.
- A Lissajous ellipse gives phase magnitude with sign ambiguity.
- Count tangent contacts with a defined axis convention.
- Match probe and scope RC time constants.
- The 0.35 rise-time rule depends on frequency-response shape.
- Sample rate, record length and analogue bandwidth are separate limits.