Chapter 4

Comparators

A comparator measures deviation from a master over a restricted range. It becomes an absolute-size measurement only after adding the reference size. High amplification trades measuring range for readable output; it does not remove master or setup errors.

4.1 Calibration and sensitivity

Set zero using a master, then replace it with the part under the same contact and support conditions. Two masters can establish local sensitivity and offset. Check the output sign before interpreting the deviation.

4.2 Mechanical, optical and electrical comparators

TypeConversion chainStrength and limitation
Mechanical dialPlunger → rack/gears → pointerSimple; friction, backlash and wear affect return
Lever or reed mechanismSmall displacement → lever rotation or elastic motionCompact amplification; limited travel
Optical/mechanical-opticalContact displacement → mirror rotation → light-spot motionLarge readable gain; alignment and vibration matter
Electrical, including LVDTDisplacement → electrical signal → displayLow moving mass, remote output; needs excitation and calibration
PneumaticGap → flow or pressure → indicationLow measuring force; sensitive to air supply and nozzle condition

An LVDT has two secondary coils connected in series opposition. At electrical null their induced voltages balance; displacement changes the differential magnitude, and phase relative to excitation distinguishes direction. Use its linear central range. For circuit details, see the Electrical and Electronic Measurements handbook.

4.3 Pneumatic gauging

In a back-pressure gauge, air passes a restriction and escapes between a nozzle and the workpiece. A smaller exit gap restricts escape and raises back pressure. In a flow-type gauge, the change in flow is measured instead. Master settings are essential because the full pressure–gap curve is nonlinear; this is also the operating distinction described in the air-gauge manufacturer's guide.

Air plugs can inspect bores using opposed jets. They average over finite jet areas and do not reproduce every local form defect. Keep supply pressure regulated and air clean; a blocked nozzle can imitate a small gap. No universal pressure–gap constant applies to all nozzle/restriction designs.

Last-minute recall

  • A comparator reports departure from a master.
  • Calibrate both zero and sensitivity; retain the sensitivity sign.
  • Cascaded displacement gains multiply.
  • Optical levers contain the mirror-reflection factor of two.
  • LVDT phase distinguishes the two sides of null.
  • Smaller air-gauge escape gap means higher back pressure.
  • Greater amplification does not guarantee lower measurement uncertainty.

Revise offline: the whole Metrology and Inspection handbook as one PDF

45 pages · 2.0 MBFree to read online. Sign in to download the PDF.We save the handbook, version and request time in My downloads. Privacy policy