| Basic and limiting sizes | B, Dmin, Dmax, dmin, dmax | mm | Capital D hole, lowercase d shaft; B common basic size. |
| Deviations and tolerance widths | ES, EI, es, ei, TH, Ts, TF, T | mm or μm | Upper/lower hole and shaft deviations; T denotes tolerance only in Chapter 1. |
| Fit quantities | Cmin, Cmax, A, Imin, Imax | mm or μm | Signed clearance and allowance are hole minus shaft; interference reverses the subtraction. |
| Tolerance factor and step | D1, D2, Dg, i, k | mm; μm for i; 1 for k | Step endpoints/geometric mean, standard factor and multiplier; local D1/D2 also taper diameters. |
| Indication and correction | xind, xref, xcorr, ex, cx | mm | Error = indication − reference; correction = −error. |
| Scale reading | sm, sv, n, N, k, LCv, LCμ, xmain, ℓs | mm; counts for n,N,k | Division lengths/counts, least counts, main reading and screw lead in mm/rev. |
| Block lengths and expansion | Li, Lstack, L20, L(T), αL | mm; K⁻¹ for αL | T is temperature in °C in the thermal section; reference 20 °C. |
| Length projections and offset | L, Lproj, a, eA, θ | mm; rad for θ | a is perpendicular Abbe offset; sign of eA depends on geometry. |
| Thread measurement | Ep, Mw, dw, dw,best, p, β, rw | mm; angle for β | E_p pitch diameter; p pitch; 2β included angle; r_w wire radius. |
| Sine-bar quantities | Lb, h, h1, h2, θ | mm; rad or degree | L_b centre spacing; h=h2−h1; differential θ in rad. |
| Taper and roller quantities | α, ℓ, H, H0, Ha, Hb, Da, Db, r | angle; mm | α half-angle, ell axial spacing, H roller top height, H0 apex datum. |
| Optical angle and level | xf, fo, sb, Rv, Ls, Δh, q, θ | mm; q in mm/m per division; rad | f_o focal length, s_b vial arc displacement, R_v curvature radius, L_s base length. |
| Reversal readings | r1, r2, b | rad, arcsecond or common angle unit | b zero offset; r2 reverses surface slope, not offset. |
| Comparator calibration | x, x0, x1, x2, y, y0, y1, y2, S | input mm; output divisions or electrical unit | S output/input, signed; local y also roughness ordinate in Chapter 6. |
| Mechanical/optical gain | Mℓ, Mj, Mtrain, li, lo, Lo | 1; mm for lengths | l_i/l_o lever arms, L_o mirror-to-screen distance. |
| Air-gauge quantities | dn, g, Ae, pb | m; m² for area; Pa | Nozzle diameter, gap, curtain area and back pressure. |
| Interference quantities | λ0, λ, nf, t, r, m, Nf, θw, bf | m or μm; 1 for n_f,m,N_f; angle for r,θw | Vacuum/air wavelengths; t film thickness in Chapter 5; r refracted angle; N_f fringe intervals. |
| Optical-flat comparison | af, δz, ℓf, G, ΔH | consistent length | Fringe lateral departure, surface departure, observed length, support separation and step. |
| Interferometer motion | Δx, Δϕ | m; rad | Double-pass displacement and phase change. |
| Form quantities | Fflat, Fround, TIR, ec, pi, c, n | mm or μm; 1 for unit normal | Minimum-zone widths, indicator range, eccentricity, sampled points and circle centre. |
| Profile coordinates | x, z, y, zˉ, L, Le, N | mm for traverse; μm for height; count | y is deviation from specified mean line; L_e evaluation length. |
| Roughness parameters | Ra, Rq, Rt, Rz, Rz,10, Rp, Rv, Ar, Pi, Vi, zp,i, zv,i | μm | Rz10 legacy ten-point height; other symbols use explicitly stated sampling conventions. |
| Machine alignment | ψ, Δu, Lt, D1, D2 | rad; mm | Transverse axis difference, traverse and end diameters for ideal taper. |
| Slope integration | ℓb, θj, zk, zk′, z0, zN, N | mm; rad for slopes; count | ell_b equal station/base spacing; primed height removes endpoint line. |
| Coordinate geometry | R, I, o, ppart, pmachine, pcentre, psurface, rp, s | mm for coordinates/radius/distance; 1 for matrices | R columns are part axes in machine coordinates; I identity; normal points toward probe centre. |
| Fitted plane | a, b, c, c1, c2, X, z, δi, s12 | a,b dimensionless; intercept/residual/distance mm | X rows contain x,y,1 with matching units; vertical regression distinguished from orthogonal fit. |
| Assembly loop | Xi, Y, ai, xi0, Y0, τi, τY, Ti | mm; 1 for a_i | tau denotes tolerance half-width here, not stress; T_i full width. |
| Statistical assembly | μi, μY, σi, σY, Cov, k, τY,RSS | mm; covariance mm²; k dimensionless | mu/sigma mean/standard deviation here, not viscosity/stress; common coverage k. |
| Selective assembly | n, j, Dmin,j, Dmax,j, dmin,j, dmax,j, Cmin,j, Cmax,j | counts; mm for sizes | n number of contributors; j permitted bin-pair index. |
| Ultrasonic quantities | su, cu, Δt, zu, xu, θu, λu, fu | m; m/s; s; rad; Hz as appropriate | One-way path, wave speed, round-trip time, depth/offset, angle from normal, wavelength and frequency. |
| Acoustic impedance and reflection | Zu, Z1, Z2, ρ, RI | kg/(m²·s); kg/m³ for density; 1 for fraction | Intensity reflection at normal incidence with no mode conversion. |
| Radiographic quantities | I0, I, μX, ℓ, ℓ1/2 | intensity; m⁻¹; m | I here intensity, not interference or identity matrix; narrow-beam attenuation. |