| Instantaneous voltage/current; DC or RMS magnitudes | v,i;V,I | V; A | AC phasors are RMS; underlining marks complex phasors where needed |
| Resistance, impedance, inductance, mutual inductance, capacitance | R,Z,L,M,C | Ω; Ω; H; H; F | Arm subscripts and physical connections are stated locally |
| Frequency, angular frequency, period, time | f,ω,T,t | Hz; rad/s; s; s | ω=2πf; T means waveform period here |
| Active/reactive/complex power and energy | P,Q,S,E | W; var; VA; J or kWh | S=P+jQ where complex power is used; Q>0 inductive |
| Measured, indicated and reference values | xi,xind,xref,y | Measurand units | Generic scalar measurements |
| Signed error, correction, limiting bound | e,c,δx,Δx,ϵ | Measurand units; fractional ϵ | Indication minus reference; Δx≥0 |
| Full scale, range, maximum display | XFS,Xrange,Xmax | Measurand units | Range endpoint convention explicit |
| Sensitivity, uncertainty, scatter | ci,ui,uy,uxˉ,sx,U,ku | [y]/[xi]; input/output units; dimensionless ku | ci is a derivative; U=kuuy |
| Mean, covariance, repeated count | xˉ,Cov(xi,xj),n | Input units; product units; count | sx uses n−1 |
| Nonlinear-model parameters | Is,a,V | A; V; V | a is effective exponential voltage in 1.4 |
| Decibel gain/loss | GdB,LdB | dB | Positive attenuation loss |
| Pointer angle and torques | θ,Td,Tc,Tb,kθ,kg | rad; N·m; N·m/rad for kθ | Controlling and braking torques oppose motion/deflection |
| Magnetic field, coil area and turns | B,Ac,N | T; m²; turns | B separately denotes an imaginary current component in 6.2 |
| Movement, shunt and multiplier | Ifs,Rm,Rsh,Rmult,RA,RV,SV,mI,mV | A; Ω; Ω/V for SV; dimensionless ratios | Full-scale current; total ammeter/voltmeter input resistance |
| Meter inductances and extension capacitances | Lm,Lsh,Cm,Cs | H; F | Movement/shunt inductance; electrostatic range capacitors |
| Loading and ohmmeter parameters | VTh,RTh,Ro,Eb,Itest,Vsense | V; Ω; A as appropriate | Thevenin source; internal ohmmeter resistance/battery |
| Form and crest factors | Kf,Kc | 1 | Kc is a crest factor only in 2.4; transformer Kc is defined in 6.2 |
| Bridge arms and detector | R1…R4,Z1…Z4,Rg,Ig,Sg,VBD,Vs | Ω; A; rad/A; V | Detector positive B to D; see each topology |
| Kelvin ratio/link resistances | P,Q,p,q,r,Rx,Rs | Ω | Local resistor labels, not power; source A–D |
| Potentiometer standard and lengths | Es,Ex,E1,E2,ls,lx,l1,l2,kp | V; m; V/m | kp is voltage gradient; working current fixed |
| Cell/load and AC potentiometer components | rcell,RL,loc,lload,Vd,Vq | Ω; m; V | Vd,Vq signed in-phase/quadrature voltages in 3.5 |
| Coil/capacitor loss factors | QL,QC,D,δloss | 1; rad | D=tanδloss |
| Capacitor equivalent and standard | Rp,Cp,Cx,Cs | Ω; F | Series-to-parallel conversion at stated frequency; probe parameters separately defined |
| Voltage-current phase and pressure-circuit lag | ϕ,ϕh,β | rad or explicit degrees | Positive ϕ is lagging current |
| Wattmeter quantities and harmonic components | W,Wa,Wb,Wc,Wq,Vpc,Icc,Vh,Ih | W; V; A | Coil polarities define sign; h=0 is DC |
| Pressure/current-circuit parameters | Rpc,Lpc,Rcc,Vr,Ir,d,dfs | Ω; H; V; A; divisions | Scale d is not CRT separation |
| Line/phase and three-meter readings | Vline,Iline,Vs,VR,Vx,Is,IR,Ix | V; A | Each method defines the measured branch |
| Energy indication and constants | NE,KE,Kh,nrpm,ϵE | count; count/kWh; Wh/rev; rev/min; 1 | Count may mean pulses or disc revolutions |
| Induction-meter fluxes and speed | Φv,Φi,Φb,Ωd,α,kd,kb | Wb; rad/s; rad; construction-dependent torque constants | α is voltage-to-pressure-flux lag |
| Transformer turns, burden and flux | Np,Ns,nc,nv,Sb,Zb,Es,Φmax | turns; 1; VA; Ω; V; Wb | CT nc=Ns/Np; PT nv=Np/Ns |
| Actual/nominal ratios, phase errors | Rc,Rv,Kc,Kv,ϵr,βc,βv,RCF | 1; radians for phase | Secondary lead positive after polarity correction |
| Excitation and transformer drops | I0,Iw,Iμ,A,B,δ,Zp,Zs,ΔV | A; rad; Ω; V | A,B real/imaginary primary-current components in 6.2 |
| Digit/count specifications | d,Nmax,Nspan,q,a,b,c,eq,uq | digits; count; input units; percentage numbers | q least count; a,b,c accuracy coefficients in 7.2 |
| DVM integration and reference | T1,T2,N1,N2,Vref,fn,H | s; count; V; Hz; 1 | H is boxcar gain at interference frequency |
| Counter gate and timebase | Tg,N,Nc,m,fclk,fx,Tx,ϵt | s; count; Hz; 1 | m whole input cycles; one-count bound stated |
| Crossing delay and slope | Δt,Δterr,Δttrig,sv | s; V/s | Corresponding same-slope crossings |
| Oscilloscope scales | ap,Sy,St,ny,nt,Vpp | 1; V/div; s/div; divisions; V | Probe multiplier applied once |
| XY amplitudes, intercepts and phase | A,B,x,y,x0,y0,ψ,NH,NV | Display units; rad; counts | ψ is Y minus X phase; tangent convention in 8.3 |
| Probe and rise-time parameters | Rp,Ri,Cp,Ci,fBW,tr | Ω; F; Hz; s | Compensated RC divider; 10–90% rise time |
| CRT and acquisition | Va,Vd,l,d,L,Le,y,Sd,fs,Ns,Δts,Trecord | V; m; m/V; Hz; count; s | L measured from plate centre; Ns samples in 8.5, turns in 6.1 |